Description
Output
- Two-wire 4-20 mA is user-selectable for linear or square root output. Digital signal based on HART
Range
- 051S_CA: 0-0.167 to 0-4000 psia
- 3051S_CD: 0-0.01 in H2O to 0-2000 psid
- 3051S_CG: 0-8.3 in H2O to 2000 psig
Accuracy
3051S_CD/CG, Ranges 2-4:
- Classic: +0.035% of span, For spans less than 10:1: ±[0.015 + 0.005(URL / Span)]% of span
- Ultra: +0.025% of span, For spans less than 10:1: ±[0.005 + 0.0035(URL / Span)]% of span
- Ultra for Flow(3051S_CD Only, Ranges 2-3): +0.04% of reading up to 8:1 DP turndown from URL, or ±[0.04 + 0.0023(URL / Reading)]% of reading to 200:1 DP turndown from URL
3051S_CD/CG, Range 5:
- Classic: +0.065% of span, For spans less than 10:1, ±[0.015 + 0.005(URL / Span)]% of span
- Ultra: +0.05% of span, For spans less than 10:1, ±[0.005 + 0.0045(URL / Span)]% of span
3051S_CD/CG, Range 1:
- Classic: +0.10% of span, For spans less than 15:1, ±[0.025 + 0.005(URL / Span)]% of span
- Ultra: +0.09% of span, For spans less than 15:1, ±[0.015 + 0.005(URL / Span)]% of span
3051S_CD, Range 0:
- Classic: +0.10% of span, For spans less than 2:1: ±0.05% of URL
- Ultra: +0.09% of span, For spans less than 2:1: ±0.045% of URL
3051S_CA, Ranges 1-4:
- Classic: +0.035% of span, For spans less than 10:1: ±[0.0065(URL / Span)]% of span
- Ultra: +0.025% of span, For spans less than 10:1: ±[0.004(URL / Span)]% of span
3051S_CA, Range 0:
- Classic: +0.075% of span, For spans less than 5:1: ±[0.025 + 0.01(URL / Span)]% of span
- Ultra: +0.075% of span, For spans less than 5:1: ±[0.025 + 0.01(URL / Span)]% of span
Power
- 3051S: 10.5-42.4 V dc with no load
- 3051S with Advanced HART Diagnostic Suite: 12-42.4 V dc with no load
Process Wetted Material
- 316L SST, Alloy C-276, Alloy 400, Tantalum, Gold-plated Alloy 400, Gold-plated 316L SST
Materials
- Sensor Wetted Parts: 316L SST, Hastelloy, Monel, Tantalum, Gold-plated SS, Gold-plated Monel
- Sensor Fill Fluid: Silicone, Inert (Halocarbon)
- Coplanar and Traditional Flanges and Adapters: Cadmium and Nickel-plated CS, 316 SST, Hastelloy, Monel
Process Flange
- Carbon Steel, SST
Features
- Ultra, ultra for flow, and classic performance
- Gauge, differential or absolute measurements
- Remote Display and Interface options
- Zero and span adjustment
- PlantWeb alerts
- Burst pressure limit: 10,000 psig (689.5 bar)
- Best-in-class, scalable architecture enhances performance, functionality and process connections
- Transmitter arrives fully assembled and leak tested for simple installation and maintenance